Design History File (DHF) for Drug-Device Combination Products: A Complete Guide

background
user-icon 14 Oct 2026

Introduction: Why Your Combination Product Needs More Than a Collection of Documents

A drug-device combination product may contain a well-characterized medicinal product and a technically mature delivery device, yet still encounter significant development or regulatory challenges. The reason is often not the absence of individual documents. It is the lack of a coherent record demonstrating how the product was designed, evaluated, verified, validated and approved for its intended use.

Consider a prefilled syringe. The medicinal product may meet its pharmaceutical quality requirements, and the syringe may satisfy applicable device requirements. However, the complete product must also demonstrate that the container, closure, materials, delivery mechanism and user interactions work together as intended. Changes to one component may affect the performance, safety or quality of another.

This is where the Design History File (DHF) becomes important.

A well-managed DHF connects user needs, design inputs, design outputs, risk management, verification, validation and design changes. It helps an organization demonstrate that development decisions were made systematically and supported by appropriate evidence.

For combination products, this documentation must also reflect the relationship between the medicinal product and the delivery system. This guide explains the purpose of a DHF, the relevant regulatory expectations, the records that typically belong in the file, and practical ways to maintain it throughout the product lifecycle.

1. What Is a Design History File?

A Design History File is traditionally understood in the United States as the collection of records demonstrating that a medical device was developed in accordance with the approved design plan and applicable design-control requirements.

Under ISO 13485:2016, Clause 7.3.10 addresses design and development files. The organization is required to maintain a file for each medical device type or medical device family, containing or referencing records that demonstrate conformity with the applicable design and development requirements.

Although the terms DHF and design and development file are often used interchangeably in industry discussions, their regulatory context matters. Organizations should identify the terminology and requirements applicable to their markets and quality management system.

For a drug-device combination product, the file should provide a clear development history, including:

  • What the product was intended to achieve.
  • Which user, patient and regulatory needs informed the design.
  • How design inputs and outputs were established and controlled.
  • Which risks were identified and how they were addressed.
  • How verification and validation demonstrated acceptable performance.
  • How design changes were assessed, approved and documented.
  • How the final design was transferred into production or manufacturing controls.

The DHF is not simply an archive of documents. It should tell a coherent, evidence-based story of product development.

2. Regulatory Expectations: FDA QMSR and the European Union

A DHF strategy should begin with the intended markets, product classification and applicable regulatory pathway. There is no single regulatory route for every drug-device combination.

United States: FDA Quality Management System Regulation

The FDA’s Quality Management System Regulation (QMSR) became effective on February 2, 2026. It amended 21 CFR Part 820 and incorporates ISO 13485:2016 by reference, aligning the US medical device quality management framework more closely with the international standard.

For organizations maintaining design-control records, this makes it important to assess their documentation against the applicable QMSR requirements and ISO 13485 provisions. The design and development file requirements in ISO 13485:2016 Clause 7.3.10 are particularly relevant.

The practical objective is to maintain records that demonstrate the design process was controlled, risks were evaluated, and the resulting product meets its specified requirements.

Organizations should not assume that existing DHF documents automatically satisfy current requirements. A documented gap assessment can identify records that need clarification, additional evidence or improved traceability.

Official reference: FDA — Quality Management System Regulation (QMSR).

European Union: Medical Device and Medicinal Product Requirements

In the EU, the applicable framework depends on how the medicinal product and device are combined, supplied and intended to be used.

Relevant considerations may include Regulation (EU) 2017/745 on medical devices (MDR), medicinal product legislation, and applicable EMA guidance. For certain integral drug-device combinations, Article 117 of the MDR is relevant to the documentation supporting device compliance in a medicinal product marketing authorisation application.

The EMA’s Quality documentation for medicinal products when used with a medical device guideline addresses product-specific quality aspects of integral, co-packaged and separately obtained devices that may affect the quality, safety or efficacy of a medicinal product.

Consequently, a DHF or design and development file should support the broader regulatory evidence strategy without being confused with the entire EU technical documentation or medicinal product dossier.

Official references:

3. DHF, Technical Documentation, DMR and DHR: Understanding the Differences

These records serve different purposes and should not be treated as interchangeable.

Record or filePrimary purpose
Design History File / design and development fileDemonstrates the history and control of product design and development.
Medical device technical documentationSupports applicable regulatory conformity assessments and demonstrates compliance with relevant requirements.
Device Master Record (DMR)Traditional US terminology for records containing production specifications and procedures needed to manufacture the device.
Device History Record (DHR)Traditional US terminology for records demonstrating that a particular manufactured batch, lot or unit was produced in accordance with applicable requirements.

The terminology and specific obligations depend on the applicable regulatory framework. Under the current FDA QMSR, manufacturers should assess their record structures against the revised regulation rather than relying solely on historical terminology.

For a combination product, some evidence may support more than one regulatory purpose. Nevertheless, each record should have a defined location, owner and relationship to the relevant requirement.

The DHF should not become a dumping ground for every manufacturing, quality or regulatory document. It should reference controlled records where appropriate and maintain clear links to the evidence supporting design decisions.

4. What Should a Combination Product DHF Contain?

The precise structure will depend on the product, development methodology, applicable requirements and company procedures. The following framework provides a practical starting point.

4.1 Design and Development Planning

The development plan establishes how the design process will be controlled.

Typical records include:

  • Design and development plan, including stages and milestones.
  • Defined responsibilities and approval authorities.
  • Planned design reviews.
  • Verification and validation strategy.
  • Risk management activities.
  • Interfaces between pharmaceutical, device, engineering, quality and regulatory teams.
  • Criteria for design transfer and change control.

For combination products, the plan should identify how decisions affecting both the drug and device will be coordinated. Separate workstreams may be necessary, but their interfaces must be managed.

4.2 User Needs and Intended Use

User needs describe what patients, healthcare professionals and other intended users need the product to accomplish.

For an injection device, these needs might include accurate dose delivery, acceptable usability, reliable operation and safe administration under specified conditions.

The records should establish the intended use, target users, use environment and relevant product requirements. Where applicable, they should also account for foreseeable use errors and patient-related risks.

User needs must be translated into requirements that can be evaluated. Statements such as “the device should be easy to use” are not sufficient on their own; the organization needs defined, assessable criteria.

For more detail, see TSQE’s article: Design Inputs for Combination Products: Turning User Needs Into Testable Requirements.

4.3 Design Inputs and Design Outputs

Design inputs establish the requirements against which the design will be developed and evaluated. Design outputs describe the resulting specifications, configurations and other deliverables.

Inputs may address:

  • Dose accuracy and delivery performance.
  • Materials and product-contact compatibility.
  • Functional and mechanical requirements.
  • Usability and human factors.
  • Sterility and packaging, where applicable.
  • Storage, transportation and environmental conditions.
  • Software, electronics or connectivity, where relevant.
  • Applicable safety, performance and regulatory requirements.

Outputs may include approved drawings, specifications, software requirements, bills of materials, component specifications, manufacturing instructions and inspection criteria.

Each important input should have an identifiable output and a defined method of verification. Where requirements are revised, the rationale and approval should be documented.

4.4 Risk Management Records

Risk management must be integrated into design and development rather than treated as an isolated activity completed near the end of the project.

For a drug-device combination product, risks may arise from device failure, drug-device interactions, incorrect dose delivery, material incompatibility, contamination, usability issues or changes to manufacturing processes.

The DHF should reference the relevant risk management records and demonstrate how identified risks informed design requirements, controls and verification or validation activities.

For example, if a delivery mechanism could cause under-delivery, the design inputs, risk controls and test protocols should collectively show how that risk was addressed and how the effectiveness of the controls was evaluated.

The key question is not merely whether a risk analysis exists. It is whether the design evidence demonstrates that the identified risks were addressed.

4.5 Design Reviews

Formal design reviews provide structured opportunities to assess whether development activities are progressing appropriately and whether significant issues remain unresolved.

Records should identify the review stage, participants, design information considered, conclusions, action items and follow-up.

For combination products, the review team may need expertise from device engineering, pharmaceutical development, quality assurance, regulatory affairs, manufacturing, clinical or human factors functions, depending on the product.

A review should result in documented decisions and actions, not simply meeting minutes stating that the project was discussed.

4.6 Design Verification and Validation

Verification and validation address different questions.

Design verification: Does the design output meet the specified design input?

Design validation: Does the resulting product meet user needs and intended uses under the applicable conditions?

Verification activities may include dimensional testing, functional testing, software testing, dose-delivery testing and material assessments, depending on the design.

Validation may involve simulated-use studies, usability validation, performance evaluation and other activities appropriate to the intended users and product risks.

Combination-product evidence may also need to address drug-device compatibility, container-closure performance, extractables and leachables, stability, transport, packaging integrity or other product-specific concerns.

Not every test belongs in every DHF. The evidence strategy should be justified by the design, risk assessment and applicable regulatory requirements.

Where a test is conducted under a separate pharmaceutical or laboratory quality system, the DHF can reference the approved report, provided that the report is identifiable, controlled and clearly linked to the relevant requirement.

4.7 Design Transfer

Design transfer establishes that the approved design can be translated into controlled manufacturing and production activities.

Relevant records may include transfer plans, approved specifications, manufacturing process requirements, acceptance criteria, training records and evidence that production arrangements are ready.

For a combination product, transfer requires attention to the interfaces between the drug, device, assembly or filling process, packaging and final product release.

The development team should be able to demonstrate that the final design is adequately defined and that the receiving manufacturing operation has the information and controls needed to produce the intended product consistently.

4.8 Approvals and Design History Records

The file should make it possible to determine which design version was approved, who authorized it, when approval occurred and which evidence supported the decision.

Drafts, obsolete specifications and unapproved changes should not be confused with the approved design baseline.

An effective document control process ensures that the current design is identifiable and that historical versions remain available when required to reconstruct earlier decisions.

5. Traceability: Connecting Requirements to Evidence

Traceability is one of the most valuable features of a well-managed DHF.

A traceability matrix can connect:

User needs → Design inputs → Design outputs → Risk controls → Verification → Validation

Not every relationship will be strictly linear, and one requirement may be supported by several records. The objective is to make the relationships clear, complete and reviewable.

Consider a prefilled syringe requirement for accurate dose delivery. The traceability record might link the user need to a measurable dose-delivery specification, the relevant design output, the associated risk controls and the approved test report demonstrating performance.

Traceability also helps identify gaps. If a requirement has no verification evidence, the team can investigate whether testing is missing, the requirement is not applicable, or the rationale needs to be documented.

The matrix should be maintained as the design evolves. An outdated matrix can create a false sense of control.

Read more in TSQE’s guide to End-to-End Traceability: From User Needs to Design Verification and Validation.

6. Managing Design Changes Throughout the Lifecycle

Design documentation does not stop at design approval. Changes may arise from supplier substitutions, manufacturing improvements, updated user feedback, component obsolescence, new risk information or regulatory developments.

A change assessment should consider the impact on:

  • Approved design inputs and outputs.
  • Product performance and safety.
  • Drug-device compatibility.
  • Risk management conclusions.
  • Verification and validation requirements.
  • Manufacturing processes and released product.
  • Labelling, regulatory submissions and approvals, where applicable.

Changes should be reviewed, verified and validated as appropriate, and approved before implementation in accordance with applicable requirements and procedures.

For example, changing a polymer component may affect mechanical performance, drug compatibility or extractables and leachables. A simple drawing revision may therefore be insufficient; the organization should document the impact assessment and determine what additional evidence is necessary.

See the supporting article: Managing Design Changes Across the Drug and Device Lifecycle.

7. Keeping the DHF Inspection-Ready

Inspection readiness should be built into daily document management rather than treated as a final preparation exercise.

A practical review should confirm that:

  1. Records are complete: Required plans, reviews, risk records and test evidence are present or appropriately referenced.
  2. Approvals are traceable: The approved design baseline and decision-makers are identifiable.
  3. Requirements are linked to evidence: Verification and validation records address the applicable requirements.
  4. Changes are controlled: The rationale, impact assessment, approvals and resulting actions are documented.
  5. Versions are clear: Obsolete documents cannot be mistaken for current approved records.
  6. Cross-functional interfaces are addressed: Drug, device, manufacturing and regulatory evidence is connected where necessary.
  7. Records can be retrieved: Staff can locate the correct evidence without relying on individual employees’ personal filing systems.

Under the FDA QMSR, manufacturers should also be prepared for inspections conducted under the updated inspection framework and for requests to review applicable quality management system records.

For a deeper practical approach, see How to Keep Design Documentation Inspection-Ready Throughout Product Development.

8. Common DHF Management Mistakes

Several recurring problems can weaken a combination product’s development evidence.

1. Treating the DHF as an archive. Documents may exist, but the relationships between requirements, risks, design decisions and tests are unclear.

2. Starting documentation too late. Teams attempt to reconstruct decisions after development is substantially complete, increasing the risk of missing rationale or approvals.

3. Separating drug and device activities without managing interfaces. Individual workstreams may appear complete while interactions between the medicinal product and device remain inadequately assessed.

4. Maintaining incomplete traceability. Requirements lack evidence, tests do not identify the requirements they address, or changes are not reflected in the matrix.

5. Failing to assess changes across the complete product. A component change is evaluated in isolation even though it may affect delivery performance, compatibility, usability or regulatory commitments.

6. Confusing documentation with proof. A procedure or plan describes what should happen; the associated records must demonstrate what actually happened and what was concluded.

7. Using unclear document ownership. Multiple teams maintain overlapping files without a defined source of truth, leading to inconsistent versions and approval gaps.

These problems can be addressed through defined procedures, document ownership, structured reviews and periodic checks of DHF completeness.

9. A Practical Roadmap for Building and Maintaining a DHF

Organizations do not necessarily need to redesign their entire documentation system to improve DHF management. A structured approach can identify the most important gaps first.

Step 1 — Define the regulatory scope. Identify the target markets, product configuration, applicable requirements and relevant quality management system procedures.

Step 2 — Establish the file structure. Define document categories, ownership, approval requirements, naming conventions and the approach to referencing records maintained in other controlled systems.

Step 3 — Map the design process. Connect planning, user needs, inputs, outputs, risk management, reviews, verification, validation and design transfer.

Step 4 — Build traceability. Establish links between requirements, risk controls and supporting evidence. Identify gaps and assign actions.

Step 5 — Assess combination-product interfaces. Confirm that the development evidence addresses relevant interactions between the medicinal product, device, packaging, users and manufacturing processes.

Step 6 — Review changes and historical records. Check whether approved changes are reflected in the current design baseline and whether earlier versions and decisions remain reconstructable.

Step 7 — Conduct a documented gap assessment. Evaluate the file against applicable requirements and internal procedures, prioritize deficiencies according to risk and regulatory significance, and assign owners and due dates.

Step 8 — Maintain readiness. Integrate DHF checks into design reviews, change control, quality management and project governance so that documentation remains current as the product evolves.

For practical implementation support, see TSQE’s DHF and Design Control service and Design Control Process service.

Conclusion & How TS Quality & Engineering Can Help

A well-managed Design History File provides more than evidence that documents have been created. It demonstrates how user needs were translated into controlled design requirements, how risks were addressed, how the product was verified and validated, and how the approved design is maintained throughout its lifecycle.

For drug-device combination products, this requires particular attention to the interfaces between the medicinal product and the delivery system. Design changes, compatibility concerns, usability and manufacturing transfer can affect the complete product and should be reflected in the development evidence.

TS Quality & Engineering supports medical device and combination-product organizations with design control, DHF development and review, traceability, risk management integration and regulatory documentation.

Our approach focuses on making development records structured, connected and useful for both product teams and regulatory assessments.

Need support with your combination product’s DHF? Visit TS Quality & Engineering’s DHF and Design Control services to discuss your documentation framework, gap assessment or design-control needs.

Regulatory note: This article provides general information and is not a substitute for product-specific regulatory assessment. Applicable requirements depend on the product configuration, intended purpose, target market and regulatory pathway.


References

  1. U.S. Food and Drug Administration (FDA). Quality Management System Regulation (QMSR).
  2. International Organization for Standardization (ISO). ISO 13485:2016 — Medical devices: Quality management systems — Requirements for regulatory purposes.
  3. European Medicines Agency (EMA). Quality documentation for medicinal products when used with a medical device.
  4. European Commission. MDCG guidance and other guidance documents for medical devices.

About the author: