A drug-device combination product must be developed with evidence that its design meets user needs, addresses relevant risks and performs as intended. When the product reaches regulatory review or an inspection, the organization should be able to explain not only what the final product looks like, but also how its requirements were established, how design decisions were made and how the resulting product was verified and validated.
This is where the Design History File (DHF) concept becomes useful. For combination products, the challenge is to maintain connected records covering device engineering, pharmaceutical development, human factors, risk management and manufacturing.
Regulatory terminology has changed in the United States, but the underlying need for design and development evidence has not disappeared. In Europe, the applicable documentation depends on the product configuration and regulatory pathway.
What Is a Design History File?
Traditionally, a DHF is the collection of records demonstrating that a medical device was developed in accordance with an approved design plan and applicable requirements.
For a drug-device combination product, it can provide a structured way to organize evidence relating to the device constituent part and its interaction with the medicinal product. A prefilled syringe, for example, may require evidence covering dose delivery, materials compatibility, usability, container-closure integrity, stability, manufacturing and labeling.
A DHF is not simply a folder of documents. Its value lies in demonstrating a logical connection between user needs, design inputs, risk controls, design outputs, verification, validation and design transfer.
What changed under FDA’s QMSR?
The FDA’s Quality Management System Regulation (QMSR) became effective on February 2, 2026, incorporating ISO 13485:2016 by reference. The regulation no longer uses the specific term Design History File in the same way as the former Quality System Regulation. Instead, the relevant design and development requirements are addressed through ISO 13485:2016 Clause 7.3 and its subclauses, including Clause 7.3.10 for the design and development file.
This is a change in regulatory terminology and structure—not permission to stop maintaining design records. Manufacturers should continue to maintain an organized, inspection-ready set of records demonstrating that design and development requirements have been met. FDA investigators may review quality-system records created before the QMSR effective date as well.
In practice, organizations may continue to use the term DHF internally, provided their documentation system satisfies the applicable requirements.
How to Build the File
1. Define the Product and Regulatory Strategy
Begin with a clear description of the product, its intended purpose, intended users, use environment and configuration.
The regulatory strategy should identify the requirements applicable to the intended markets. In the EU, this includes assessing the relevant medicinal-product legislation and Medical Device Regulation (EU) 2017/745 (MDR), including Article 117 where applicable.
The pathway depends on how the device and medicinal product are combined and supplied. An integral prefilled syringe may raise different conformity-assessment questions from a separately supplied delivery device. Not every combination product requires the same submission or device conformity-assessment route.
Document the rationale for the selected pathway, applicable requirements, regulatory responsibilities and planned evidence.
2. Establish a Design and Development Plan
The plan defines how development will be controlled. It should identify:
- Development stages, responsibilities and resources.
- Planned design reviews and approval criteria.
- Verification, validation and risk-management activities.
- Required input and output records.
- Design transfer, change control and record-retention arrangements.
Combination products often involve pharmaceutical development, device engineering, quality assurance, regulatory affairs, human factors and external suppliers. Define how these teams will coordinate and approve decisions.
3. Document Design Inputs
Design inputs translate user needs, intended use and applicable requirements into clear, testable requirements.
For a prefilled syringe, inputs might include dose accuracy, delivery force, compatibility with the drug formulation, protection against leakage, usability requirements, sterility-related requirements and performance during storage and transport.
Inputs should also address relevant regulatory, safety, performance, packaging and labeling requirements. Where the medicinal product may interact with device materials, compatibility and potential effects on product quality should be considered.
Avoid vague statements such as “the syringe must be easy to use.” Define how usability will be assessed and what constitutes an acceptable result.
4. Maintain Design Outputs and Risk Management Records
Design outputs describe what will be manufactured and how the design requirements will be met. Depending on the product, these may include drawings, specifications, material requirements, software specifications, packaging designs, labeling and manufacturing instructions.
Risk-management records should connect identified hazards with risk controls and evidence that those controls work. Consider risks arising from the combined product, not only from the drug or device in isolation.
For example, a delivery mechanism may function correctly during a bench test but still create a dosing risk if users misunderstand the instructions. The file should link the relevant hazard, design control, verification evidence and usability validation.
5. Record Design Reviews, Verification and Validation
These activities serve different purposes and should not be treated as interchangeable.
- Design reviews: Document multidisciplinary evaluation of design progress, issues, decisions and required actions.
- Design verification: Demonstrate that design outputs meet design inputs—for example, testing whether a syringe delivers the specified dose.
- Design validation: Demonstrate that the resulting product meets user needs and intended uses under defined conditions, using appropriate methods and representative products.
For combination products, validation may need to consider the complete system, including the medicinal product, delivery device, packaging, labeling and intended user interactions. The exact evidence depends on the product and its risks.
Retain protocols, acceptance criteria, results, deviations, conclusions and approvals. A test report without a clear link to the requirement it evaluates is less useful during an inspection.
6. Control Design Changes and Transfer
Design changes should be assessed before implementation to determine their impact on performance, safety, risk controls, regulatory submissions, manufacturing and previously completed testing.
For example, changing a syringe component material may affect drug compatibility, delivery performance or extractables and leachables assessments. The organization should document the rationale for any additional testing and the approvals required.
Design transfer records should demonstrate that the approved design has been translated into production specifications and processes. This may include manufacturing instructions, supplier requirements, process validation where applicable, inspection methods and acceptance criteria.
7. Link the Evidence Through Traceability
A practical file should allow reviewers to follow the development logic without repeatedly searching across disconnected systems.
A traceability matrix can connect:
User need → Design input → Risk control → Design output → Verification/validation evidence → Approval
This helps reveal missing tests, unverified requirements and risk controls without adequate supporting evidence. It also makes the impact of design changes easier to assess.
8. Keep the File Inspection-Ready
The design and development file should contain or reference the records needed to demonstrate compliance. Maintain document identifiers, revision status, approval records, test reports, meeting decisions and links to controlled records.
For FDA-regulated devices subject to design and development requirements, use ISO 13485:2016 Clause 7.3 as the core reference under QMSR. For EU products, ensure the file aligns with the applicable MDR or medicinal-product requirements and supporting technical documentation.
The Medzus white paper on prefilled syringes provides a useful example of organizing evidence across planning, inputs, outputs, verification, validation, design reviews, transfer and post-market activities. Its structure can inform a file index, but the contents should be tailored to the product’s actual regulatory pathway and risks.
Conclusion & How TS Quality & Engineering Can Help
A well-structured design history file—or design and development file under the current FDA framework—helps demonstrate that a combination product has been developed systematically, with clear links between user needs, risks, design decisions and supporting evidence. The objective is not to create more paperwork; it is to make the development rationale understandable, traceable and defensible.
TS Quality & Engineering (TSQE) supports medical device and combination-product organizations with design controls, risk management, technical documentation and quality-system activities. We can help review existing development records, identify documentation gaps and establish a practical structure aligned with applicable regulatory requirements.
References
- U.S. Food and Drug Administration (FDA). Quality Management System Regulation (QMSR). FDA website.
- U.S. Food and Drug Administration (FDA). QMSR – Design and Development. FDA presentation.
- European Medicines Agency (EMA). Questions and answers on implementation of the Medical Devices and In Vitro Diagnostic Medical Devices Regulations. EMA guidance.
- European Medicines Agency (EMA). Quality documentation for medicinal products when used with a medical device. EMA guidance.
- Medzus. Design History File (DHF) & Regulatory Requirements – Prefilled Syringes as Drug-Device Combination Product. White paper supplied as a reference.