During an FDA inspection or an EU regulatory assessment, having the required design documents is only part of the challenge. The organization must also demonstrate that the records are controlled, complete, consistent with the approved product design and readily available to support important development decisions.
For a drug-device combination product, this becomes more complex because evidence may be distributed across pharmaceutical development, device engineering, risk management, human factors, manufacturing and regulatory affairs. A test report may be approved while the associated design specification has changed. A risk assessment may identify a control that cannot be traced to verification evidence. Or manufacturing may use a revised drawing while an older version remains in a related record.
These situations can make it difficult to demonstrate that the product was developed and maintained under controlled conditions.
Inspection readiness should therefore be established throughout development, not treated as a document-cleanup exercise before an inspection. A practical documentation process helps the organization maintain traceability, reduce avoidable findings and explain the rationale behind its design decisions.
What Does Inspection-Ready Design Documentation Mean?
Inspection-ready documentation is a controlled set of records that allows an organization to demonstrate how design requirements were established, how risks were addressed, how the design was verified and validated, and how the approved design was transferred and maintained.
In the United States, the FDA’s Quality Management System Regulation (QMSR), effective February 2, 2026, incorporates ISO 13485:2016 by reference. Design and development controls, including the control of design changes, remain important. The design and development file should contain or reference the records needed to demonstrate conformity with applicable requirements.
For combination products, the documentation must also reflect the applicable drug and device regulatory pathway. In the EU, the required evidence depends on the product configuration and relevant medicinal-product and medical-device requirements. One documentation structure may be useful across markets, but it should not be assumed that every product follows the same regulatory route.
The practical objective is straightforward: a reviewer should be able to locate the relevant record, understand its status and follow its connection to the product requirement or decision it supports.
How to Keep Documentation Inspection-Ready
1. Maintain an Approved Design and Development Plan
The design and development plan should serve as the starting point for organizing the records. It should identify the development stages, responsible functions, planned reviews and required verification and validation activities.
As development progresses, the plan should be reviewed and updated when necessary. If a new compatibility study or additional human-factors assessment becomes necessary, the plan and associated records should reflect the decision.
Practical check: Does the plan provide a reliable roadmap to the records supporting each development stage? RQM+ recommends treating the design history as a continuing record rather than assembling it only when an inspection is announced.
2. Apply Document Control Consistently
Each controlled record should be identifiable, approved where required and linked to the correct product or design revision.
Document control should address document identification, revision history, approval status, dates, responsibilities and access to the authoritative version. Superseded documents should remain retrievable when needed for historical review, but they must not be mistaken for current approved instructions.
For example, a verification report may refer to drawing revision C, while manufacturing has already released revision D. The organization should be able to establish whether the report remains applicable and whether the change required additional testing.
Practical check: Can staff identify the current approved version of a drawing, specification, test protocol or instruction without relying on personal email folders?
3. Maintain Traceability Between Requirements and Evidence
A traceability matrix is one of the most useful tools for making design documentation inspection-ready. It should connect user needs and design inputs with outputs, risk controls, verification results and validation evidence.
For a combination product, the matrix should also address relevant interactions between the medicinal product and the delivery device.
Consider a requirement that an autoinjector must deliver the intended dose. The supporting evidence may include a design specification, a performance test report, risk-control references and usability evidence where relevant. A reviewer should be able to navigate these records without reconstructing the relationships manually.
Traceability should be updated when requirements, design outputs or test evidence change. A matrix created only at design freeze may conceal gaps that developed earlier in the project.
4. Keep Design Outputs Consistent With the Released Product
Design outputs include the drawings, specifications, bills of materials, software requirements, packaging information and labeling that define the product and support its manufacture. They should be reviewed and approved before release, as applicable.
For combination products, consistency across records is particularly important. A component specification, supplier material description and drug compatibility report should not refer to conflicting materials or product versions.
A practical approach is to conduct periodic consistency checks across the design file, manufacturing specifications and relevant regulatory documentation. This is especially useful before design transfer, process changes and submission milestones.
5. Record Design Reviews and Close Actions
A design review should produce more than meeting minutes. The record should identify the design stage reviewed, participants, issues raised, decisions made, actions assigned and evidence of closure.
Open actions should have an owner and a defined status. If a design review identifies a potential risk involving incorrect dose delivery, the file should show how the issue was evaluated, what action was taken and whether additional testing or a design change followed.
An inspector may ask not only whether a review occurred, but also how the organization resolved issues identified during that review.
Practical check: Can each significant action be traced from identification to a documented decision and closure?
6. Control Changes and Preserve the Development History
Design documentation must remain reliable when the product changes. A supplier change, software update, component modification or revised instruction for use should trigger an assessment of the affected requirements, risks, testing and regulatory obligations.
The organization should preserve the evidence supporting the earlier approved design while maintaining clear identification of the current version. It should not overwrite historical records in a way that removes the rationale for earlier decisions.
For EU products, significant-change assessments must be based on the applicable regulatory provisions. MDCG guidance on significant changes under MDR Article 120 is specifically relevant to certain legacy devices and should not be applied as a universal test for all products.
7. Conduct Periodic Readiness Reviews
Inspection readiness should be checked at defined milestones and periodically during development. A short internal review can identify missing approvals, incomplete traceability, unresolved actions, obsolete documents or evidence that cannot be located promptly.
For combination products, include representatives from quality, regulatory affairs, engineering, pharmaceutical development and manufacturing where relevant. Each function may hold records needed to explain the complete product.
A useful readiness exercise is to select one design input and trace it through the risk controls, design outputs, test results and approvals. Repeat the exercise for a design change and a manufacturing transfer. Gaps become visible before an external reviewer identifies them.
Conclusion & How TS Quality & Engineering Can Help
Inspection-ready design documentation is the result of consistent control throughout product development, not a last-minute administrative exercise. Approved records, clear traceability, controlled design outputs and documented decisions help demonstrate that the combination product has been developed systematically and that its risks have been addressed.
For drug-device combination products, the additional challenge is maintaining consistency across device engineering, pharmaceutical quality, manufacturing and regulatory requirements. A well-organized design and development file helps the organization respond to inspections more efficiently while supporting sound decisions throughout the product lifecycle.
TS Quality & Engineering (TSQE) supports medical-device and combination-product organizations with design controls, risk management, technical documentation and quality-system implementation. We can help review existing documentation, assess traceability gaps and establish practical processes to maintain inspection readiness.
Related reading
Link these three previously developed TSQE articles within the published version:
- How to Build a Design History File for a Drug-Device Combination Product
- What Belongs in a Combination Product DHF? A Practical Documentation Framework
- Managing Design Changes Across the Drug and Device Lifecycle
References
- U.S. Food and Drug Administration (FDA). Quality Management System Regulation (QMSR). Official overview of the regulation effective February 2, 2026, including incorporation of ISO 13485:2016 by reference. FDA QMSR.
- U.S. Food and Drug Administration (FDA). QMSR: Risk and Design and Development. FDA presentation explaining design and development and risk management under the updated framework. FDA presentation (PDF).
- International Organization for Standardization (ISO). ISO 13485:2016 — Medical devices — Quality management systems — Requirements for regulatory purposes. Relevant provisions include Clause 4.2 on documentation, Clause 7.3 on design and development, and Clause 7.5 on production and service provision. ISO standard page.
- European Parliament and Council. Regulation (EU) 2017/745 on medical devices (MDR). See Annex II for technical documentation and Annex IX for conformity assessment based on a quality management system and technical documentation. Official EUR-Lex text.
- European Commission — Medical Device Coordination Group (MDCG). Guidance: MDCG endorsed documents and other guidance. Official directory of guidance for applying the MDR and IVDR, including relevant technical documentation and regulatory implementation guidance. European Commission guidance directory.
- U.S. Food and Drug Administration (FDA). Quality Management System Regulation — Frequently Asked Questions. Additional clarification on the QMSR framework and implementation. FDA QMSR FAQs.