A Design History File (DHF) can contain hundreds of pages and still fail to tell a clear development story. Test reports may be available, risk assessments completed and design reviews approved, yet important connections between these records may be missing.
For a drug-device combination product, this challenge is particularly significant. Evidence may be distributed across pharmaceutical development, device engineering, human factors, quality assurance, regulatory affairs and manufacturing. The practical question is not simply which documents to collect, but how to organize them so that each important design decision can be traced to supporting evidence.
This article provides a practical documentation framework for a combination product such as a prefilled syringe, autoinjector or drug delivery system.
First, Understand What the File Needs to Demonstrate
The traditional DHF concept is associated with the US medical-device Quality System Regulation. Since the FDA’s Quality Management System Regulation (QMSR) became effective on February 2, 2026, ISO 13485:2016 has been incorporated by reference. Clause 7.3.10 addresses the design and development file, which must contain or reference records needed to demonstrate conformity with design and development requirements. The underlying need for design evidence remains.
In the EU, the applicable documentation depends on the product configuration and regulatory route. For medicinal products used with medical devices, EMA guidance addresses device-related quality information that may affect the medicinal product’s quality, safety and efficacy. Article 117 of the EU Medical Device Regulation may be relevant to certain integral drug-device combinations.
Therefore, the framework below is a practical document-management structure, not a universal list of mandatory documents. Each organization should tailor it to its product, market and applicable requirements.
A Practical Documentation Framework
The following structure builds on the stages and output documents described in the Medzus white paper on prefilled syringes, while focusing on how to organize, connect and review the evidence.
1. Product Definition and Regulatory Strategy
Start with a short section that explains what is being developed and why.
Include:
- Product description, intended purpose and intended users.
- Product configuration and constituent parts.
- Target markets and regulatory pathway assessment.
- Applicable regulations, standards and guidance.
- Roles and responsibilities across the drug and device teams.
Practical check: Can a reviewer understand the product configuration and regulatory rationale without reading the entire development history?
For example, identify whether a prefilled syringe is an integral part of the medicinal product or whether the delivery device is supplied separately. This distinction may affect the applicable regulatory documentation.
2. Design and Development Planning
This section establishes how the development process will be controlled.
Include:
- Approved design and development plan.
- Development stages, milestones and approval gates.
- Responsibilities and required specialist input.
- Planned design reviews, verification and validation.
- Risk-management planning and traceability arrangements.
- Change-control and design-transfer procedures.
Keep the plan current as development progresses. If a usability study or additional compatibility assessment becomes necessary, the file should show how the activity was incorporated into the development process.
Practical check: Do the planned activities, responsibilities and acceptance criteria provide a clear route from initial requirements to design release?
3. User Needs and Design Inputs
This section should explain what the product must achieve, including its performance, safety and usability requirements.
Include:
- User needs and intended-use requirements.
- Device performance and functional specifications.
- Relevant medicinal-product characteristics.
- Material compatibility requirements.
- Human-factors and labeling requirements.
- Applicable regulatory requirements and risk-control requirements.
For a prefilled syringe, relevant inputs might include dose accuracy, injection force, drug-contact material requirements, storage conditions and requirements for correct administration.
Avoid keeping user needs in one document and engineering specifications in another without a clear relationship between them.
Practical check: Can every important design input be linked to its source, approval and planned method of verification or validation?
4. Design Outputs and Controlled Specifications
Design outputs describe the approved design and provide the information needed to manufacture and inspect the product.
Include:
- Drawings, component specifications and material requirements.
- Product and device performance specifications.
- Packaging and labeling specifications.
- Instructions for use (IFU).
- Manufacturing and assembly instructions.
- Relevant acceptance criteria and inspection requirements.
For combination products, consider both the device constituent part and interfaces with the medicinal product.
For example, changing a syringe stopper material may affect not only component specifications but also drug compatibility, product quality and stability. The file should identify the approved design version and link it to the assessments supporting its suitability.
Practical check: Are the released specifications consistent with the design version that was tested and approved?
5. Risk Management and Design Reviews
Risk-management records should connect hazards, risk controls and evidence demonstrating that those controls are effective. Design reviews should document whether development remains on track and what actions are needed.
Include:
- Risk-management plan and risk assessments.
- Hazard analyses and relevant failure-mode analyses.
- Risk-control measures and verification evidence.
- Design review minutes and decisions.
- Action items, owners and closure evidence.
- Assessment of residual risks and relevant benefit-risk conclusions.
A useful example is the risk of incomplete drug delivery. The file might connect the hazard analysis to a design requirement, a test protocol, test results and the design review that accepted the evidence.
Practical check: Can the reviewer follow the risk from identification through control and evaluation, rather than finding the risk assessment and test report as unrelated documents?
6. Verification, Validation and Compatibility Evidence
This section often becomes large, so organize it by requirement, test purpose and product configuration rather than by file type alone.
Include, as applicable:
- Verification protocols, acceptance criteria and reports.
- Device performance testing.
- Drug-device compatibility assessments.
- Relevant stability, sterility and packaging evidence.
- Biocompatibility evidence where applicable.
- Usability and human-factors validation.
- Clinical or performance evaluation evidence where required.
- Deviations, investigations and approved conclusions.
Verification demonstrates that design outputs meet design inputs. Validation addresses whether the resulting product meets user needs and intended use. These activities should not be treated as interchangeable.
For instance, a device may pass mechanical delivery testing but still present a usability problem if intended users misunderstand an administration step. Both the technical performance and the relevant user interaction need appropriate evidence.
Practical check: Does each report identify the requirement being assessed, the tested product version, acceptance criteria, results and conclusion?
7. Design Transfer and Manufacturing Readiness
A design is not fully supported by documentation simply because the prototype performs well. The organization must also establish that the approved design can be transferred into controlled production.
Include:
- Design transfer records and approvals.
- Released manufacturing specifications.
- Supplier and incoming-component requirements.
- Relevant process and sterilization validation.
- In-process and finished-product controls.
- Manufacturing readiness decisions.
For a prefilled syringe, this may involve evidence relating to component assembly, filling, sterilization strategy and container-closure integrity, depending on the process and product.
Practical check: Can production identify which approved specifications and acceptance criteria apply to the product being manufactured?
8. Changes, Regulatory Evidence and Lifecycle Records
The framework should also show how design changes and post-launch information are managed.
Include or reference:
- Design change requests and impact assessments.
- Required re-verification or revalidation.
- Regulatory submission evidence and relevant correspondence.
- Approved labeling and submission versions.
- Complaint, vigilance and post-market information relevant to design.
- Corrective and preventive action (CAPA) records where applicable.
Not every post-market record needs to be duplicated in the DHF. Controlled references to the relevant quality-system records may be more effective, provided the applicable requirements are met and the records remain accessible.
Practical check: If a component changes after launch, can the organization identify the affected requirements, risks, tests, approvals and regulatory assessments?
The Most Useful Tool: A DHF Index and Traceability Matrix
A document index helps people find records. A traceability matrix helps them understand the relationships between those records. Both should be maintained.
A simple index could use the following fields:
| Field | What to record |
| Document ID | Unique identifier |
| Document title | Clear description of the record |
| DHF section | Relevant framework category |
| Design version | Product or design revision covered |
| Approval status | Draft, approved, superseded, etc. |
| Related requirement or risk | Traceability reference |
| Record location | Link to the controlled document |
| Owner | Responsible function or person |
The traceability matrix can then connect a user need to its design input, risk control, design output, verification or validation evidence, and approval.
This avoids two common problems: missing evidence that is only discovered during a review, and duplicated documents that create uncertainty about which version is authoritative.
Conclusion & How TS Quality & Engineering Can Help
An effective combination product DHF is not defined by its page count. It is defined by whether the organization can demonstrate a coherent, traceable development process and quickly locate the evidence supporting important design decisions.
TS Quality & Engineering (TSQE) supports medical-device and combination-product organizations with design controls, risk management, technical documentation and quality-system activities. We can help establish a practical file structure, review document traceability and identify gaps before a regulatory submission or inspection.
References
- U.S. Food and Drug Administration (FDA). Quality Management System Regulation (QMSR). FDA website.
- U.S. Food and Drug Administration (FDA). QMSR: Design and Development. FDA presentation.