Every medical device product (or process used to make medical devices) carries the risk of failure. To mitigate such risks, medical device manufacturers implement robust risk management tools. One of the most common tools is Failure Modes and Effects Analysis (FMEA). However, while FMEA is valuable, it is not a substitute for a comprehensive risk management system, as required by ISO 14971.
In this article, we clarify the important differences between FMEA and ISO 14971, and explain why compliance with ISO 14971 is essential for medical device manufacturers.
What is FMEA?
According to the American Society for Quality, FMEA is a stepwise approach to identify every failure in:
- Design,
- Manufacturing process
- Products, or
- Services
As a process, FMEA is based on identification of failure modes. But for medical device manufacturers, it is crucial to understand that not all stem from failures or malfunctions. Some risks may be inherent to the intended use of the device and may not present as a failure per se. While FMEA enhances product reliability, it should not be solely relied upon for medical device risk management.By proactively assessing where and how a system might fail, FMEA helps organizations increase the reliability and safety of their products. For medical devices, FMEA is especially useful during the design and development phases to evaluate components, sub-assemblies, and materials.
Understanding ISO 14971
ISO 14971 is the internationally recognized standard that sets out requirements for risk management in the medical device industry. Unlike FMEA that identifies failure modes, ISO 14971 identifies hazards (potential sources of harm) and hazardous situations. After identifying the risks, ISO 14971 also requires manufacturers to:
- Estimate severity of these harms
- Their probability of occurrence
In fact, the risk is defined here as the combination of severity of the identified harm and its probability of occurrence..
Why are FMEA and ISO 14971 often confused?
The main source of confusion that exists between ISO 14971 and FMEA is the terminology. Although they are different in their meaning, they still “sound” the same. Let’s look at them to have a better understanding.
Common terms Used in FMEA
- Failure modes
- Effects of failure
- Causes of failure
- Process controls
- Severity
- Occurrence
- Detectability
- Risk priority number
- Recommended actions
Common terms Used in ISO 14971
- Hazards
- Hazardous situations
- Severity
- Probability of occurrence
- Risks
- Risk acceptability
- Risk controls
While there may be overlap in terminology, “failure modes” are not synonymous with “hazardous situations,” and “effects of failure” differ from the concept of “harm” as defined in ISO 14971. Such coincidences can lead medical device manufacturers to use FMEA as a tool for medical device risk management. However, doing so won’t make them compliant to ISO 14971 and thus, could leave significant gaps and will not ensure compliance with the standard.
Key Elements of ISO 14971 Risk Management
What is how ISO 14971 explains risk management that makes it so unique? Well, it presents a multi-step process of risk management entailing:
- Risk management planning
- Risk analysis
- Risk evaluation
- Risk control
- Evaluation of residual risk acceptability
- Risk management review and report
- Production/post production monitoring
This systematic approach ensures a complete risk management process throughout the entire lifecycle of a medical device.
Can FMEA and ISO 14971 be used together?
Yes. Both the frameworks can be used together in the medical device industry for collective reasons. While FMEA can be helpful, as it helps build products that function as intended and rarely fail, ISO 14971 can be equally helpful in setting the organization’s approach to risk management.
In fact, Annex B of ISO 14971 mentions some techniques for supporting risk analysis:
- Failure Mode Effects Analysis (FMEA)
- Fault Tree Analysis (FTA)
- Hazard and Operability Study (HAZOP)
- Preliminary Hazard Analysis (PHA)
- Hazard Analysis and Critical Control Points(HACCP)
According to ISO TR 24971, the techniques mentioned above are “complementary, and it’s necessary to use more than one of them in order to support a thorough and complete risk analysis.”
This statement implies that the use of ISO 14971 and FMEA is overall beneficial for the medical device manufacturing organization.

This image compares two methodologies used in risk management: FMEA (Failure Modes and Effects Analysis) and ISO 14971, specifically within the context of quality and engineering management.
- Left side (FMEA):
- The image includes an icon of an alarm (typically symbolizing a warning or an issue) and a magnifying glass over a gear, indicating a focus on identifying and analyzing failure points or issues in a system.
- FMEA is a systematic method for evaluating potential failure modes in a product or process and understanding their effects to prioritize corrective actions. It’s typically used to evaluate the risk of failure before they occur.
- Right side (ISO 14971):
- The balance scale symbolizes weighing the potential benefits of addressing a risk versus the cost or challenges involved in mitigating it.
- ISO 14971 is a risk management standard specifically designed for medical devices. It helps manufacturers apply a systematic approach to identifying hazards, estimating and evaluating the associated risks, and controlling those risks to acceptable levels.
Why Choose TSQ&E for Your Risk Management Needs?
At TSQ&E, our team of regulatory and compliance experts is dedicated to guiding medical device manufacturers through the intricacies of risk management and regulatory compliance. With deep knowledge of both FMEA and ISO 14971, we ensure our clients achieve and maintain full compliance while optimizing product safety and effectiveness..